Título

True coincidence summing corrections for an extended energy range HPGe detector

Autor

MARIA ELENA MONTERO CABRERA

Nivel de Acceso

Acceso Abierto

Referencia de datos

doi: http://dx.doi.org/10.1063/1.4927193

Resumen o descripción

True coincidence summing (TCS) effect for natural radioactive families of U-238 and Th-232 represents a problem when an environmental sample with a close source-detector geometry measurement is performed. By using a certified multi-nuclide standard source to calibrate an energy extended range (XtRa) HPGe detector, it is possible to obtain an intensity spectrum slightly affected by the TCS effect with energies from 46 to 1836 keV. In this work, the equations and some other considerations required to calculate the TCS correction factor for isotopes of natural radioactive chains are described. It is projected a validation of the calibration, performed with the IAEA-CU-2006-03 samples (soil and water).

Fecha de publicación

2015

Tipo de publicación

Artículo

Versión de la publicación

Versión aceptada

Formato

application/pdf

Idioma

Inglés

Relación

&

Montero-Cabrera, M. E. (2015, July). True coincidence summing corrections for an extended energy range HPGe detector. In AIP Conference Proceedings (Vol. 1671, No. 1, p. 030004). AIP Publishing.