Design and fabrication of a customized partial hip prosthesis employing CT-Scan data and lattice porous structures
Autor
JORGE CORONA CASTUERA
Daniela Rodríguez Delgado
John Henao
Juan Carlos Castro Sandoval
Carlos Agustín Poblano Salas
Nivel de Acceso
Acceso Abierto
Identificador alterno
doi: https://doi.org/10.1021/acsomega.0c06144
eissn: 2470-1343
Materias
Anatomy - (PALABRA CLAVE DEL AUTOR) Implants - (PALABRA CLAVE DEL AUTOR) Porosity - (PALABRA CLAVE DEL AUTOR) Lattices - (PALABRA CLAVE DEL AUTOR) Chemical structure - (PALABRA CLAVE DEL AUTOR) INGENIERÍA Y TECNOLOGÍA - (CTI) CIENCIAS TECNOLÓGICAS - (CTI) TECNOLOGÍA MÉDICA - (CTI) PRÓTESIS - (CTI) PRÓTESIS - (CTI)
Resumen o descripción
As a larger elderly human population is expected worldwide in the next 30 years, the occurrence of aging-associated illnesses will also be increased. The use of prosthetic devices by this population is currently important and will be even more dramatic in the near future. Hence, the design of prosthetic devices able to reduce some of the problems associated with the use of current components, such as stress shielding, reduced mobility, infection, discomfort, etc., becomes relevant. The use of additive manufacturing (AM) and the design fabrication of self-supported cellular structures in the biomedical area have opened up important opportunities for controlling the physical and mechanical properties of hip implants, resulting in specific benefits for the patients. Different studies have reported the development of hip prosthetic designs employing AM, although there are still opportunities for improvement when it comes to customized design and tuning of the physical and mechanical properties of such implants. This work shows the design and manufacture by AM of a personalized stainless-steel partial hip implant using tomography data and self-supported triply periodic minimal surface (TPMS) cell structures; the design considers dimensional criteria established by international standards. By employing tomography data, the external dimensions of the implant were established and the bone density of a specific patient was calculated; the density and mechanical properties in compression of the implant were modulated by employing an internal gyroid-type cell structure. Using such a cell structure, the patient’s bone density was emulated; also, the mechanical properties of the implant were fine-tuned in order to make them comparable to those reported for the bone tissue replaced by the prosthesis. The implant design and manufacturing methodology developed in this work considered the clinical condition of a specific patient and can be reproduced and adjusted for different types of bone tissue qualities for specific clinical requirements.
Funding: This research was funded by the Mexican National Council of Science and Technology (CONACYT), specifically through projects CB 253365 and FOINS 297683.
Notes: The authors declare no competing financial interest.
The authors would like to thank the funding provided to this work by the Mexican National Council of Science and Technology (CONACYT), especially through FOINS 297683 and CB 253365 projects. The authors are also grateful to CONACYT’s Catedras program, project 848. CT-scan measurements were performed at PET/CT Unit, School of Medicine, UNAM, whereas mechanical testing of self-supported structures was carried out at CIDESI-CENTA; the support provided by both institutions is deeply appreciated. An IRB approval and a letter of consent were considered for the development of this research.
PMCID # PMC7970571.
Editor
American Chemical Society
Fecha de publicación
2021
Tipo de publicación
Artículo
Versión de la publicación
Versión publicada
Recurso de información
Formato
application/pdf
Fuente
ACS Omega, vol. 6, no. 10 (march), p. 6902-6913
Idioma
Inglés
Relación
&
Poblano-Salas, C. A. (2021). Design and fabrication of a customized partial hip prosthesis employing ct-scan data and lattice porous structures. ACS Omega, 6(10), 6902-6913. https://doi.org/10.1021/acsomega.0c06144
Audiencia
Público en general
Sugerencia de citación
Corona-Castuera, J., Rodriguez-Delgado, D., Henao, J., Castro-Sandoval, J. C.,
Repositorio Orígen
CIATEQ Digital
Descargas
865